How to accurately calculate the capillary length after the high and low temperature test chamber coil

<

High and low temperature test chambers need to be replaced during the repair process. We can install a higher precision pressure gauge between the original capillary and the dry filter. The return air pipe is not connected to the compressor and is open. Start the compressor of the high and low temperature test chamber. When the pressure gauge reading is stable, record the reading at this time as the basis for determining the length of the new capillary. Then stop the compressor, remove the old capillary from the pressure gauge, and replace the new capillary with about 4m. The high and low temperature test chamber compressor is started, and the steady reading of the pressure gauge is generally higher than the previously recorded reading. A new capillary is cut as appropriate, and the new capillary length is optimal when the stable reading is substantially the same as the previously recorded reading.