Manual of human α1-acid glycoprotein (α1-AGP) ELISA detection kit

<

Human α1-acid glycoprotein (α1-AGP) ELISA detection kit This reagent is for research use only Specimens: serum or plasma

Test principle:
The α1-AGP kit is a solid-phase sandwich enzyme-linked immunosorbent assay (ELISA). Standards with known concentrations of α1-AGP and samples with unknown concentrations are added to microwell enzyme plates for detection. First, α1-AGP and biotin-labeled antibody were incubated at the same time. After washing, HRP labeled with avidin was added. After incubation and washing, the unbound enzyme conjugate is removed, and then substrates A and B are added to act simultaneously with the enzyme conjugate. Produce colors. The color depth is proportional to the concentration of α1-AGP in the sample.

Kit contents and preparation kit components (stored at 2-8 ° C) 96-well configuration 48-well configuration Preparation
96/48 serving microplate 1 plate (96T) half plate (48T) ready-to-use plastic membrane plate cover 1 half-plate ready-to-use standard product: 4.0g / L 1 bottle (0.6ml) 1 bottle ( 0.3ml) Dilute the blank control according to the instructions 1 bottle (1.0ml) 1 bottle (0.5ml) ready-to-use standard dilution buffer 1 bottle (5ml) 1 bottle (2.5ml) ready-to-use biotin-labeled anti-α1 -AGP antibody 1 bottle (6ml) 1 bottle (3.0ml) ready-to-use affinity streptavidin-HRP 1 bottle (10ml) 1 bottle (5.0ml) ready-to-use wash buffer 1 bottle (20ml) 1 bottle (10ml ) Dilute according to the instructions Substrate A 1 bottle (6.0ml) 1 bottle (3.0ml) Ready-to-use substrate B 1 bottle (6.0ml) 1 bottle (3.0ml) 1 bottle of stop solution 1 (6.0ml) 1 Bottle (3.0ml) Ready-to-use

Bring your own materials
1. Distilled water.
2. Sampler: 5ul, 10ul, 50ul, 100ul, 200ul, 500ul, 1000ul.
3. Oscillator and magnetic stirrer etc.

safety
1. Avoid direct contact with the stop solution and substrates A and B. Once exposed to these liquids, rinse with water as soon as possible.
2. Do not eat, drink, smoke or use cosmetics during the experiment.
3. Do not use your mouth to absorb any ingredients in the kit.

Operation notes
1. Reagents should be stored according to the label instructions and returned to room temperature before use. The sparse standards should be discarded and cannot be stored.
2. The slats not used in the experiment should be immediately returned to the packaging bag, sealed and stored to avoid deterioration.
3. Unused other reagents should be packed or covered. Do not mix reagents of different batches. Use before warranty.
4. Use a disposable pipette tip to avoid cross-contamination, and avoid using a sampler with a metal part when drawing the stop solution and substrates A and B.
5. Use a clean plastic container to prepare the washing solution. Mix all components and samples in the kit thoroughly before use.
6. When washing the enzyme-labeled plate, it should be fully patted dry. Do not put the absorbent paper directly into the enzyme-labeled reaction well to absorb water.
7. Substrate A should evaporate to avoid opening the lid for a long time. Substrate B is sensitive to light and avoid prolonged exposure to light. Avoid contact with hands, it is toxic. The OD value should be read immediately after the experiment is completed.
8. The order of adding reagents should be the same to ensure that all wells are incubated for the same time.
9. Perform the incubation operation according to the time, the amount and order of the liquid indicated in the instructions.

Sample collection, processing and storage methods
1. Serum-Avoid any cell stimulation during the operation. Use test tubes free of pyrogens and endotoxins. After collecting blood, centrifuge at 1000 × g for 10 minutes to separate the serum and red blood cells quickly and carefully.
2. Plasma ----- EDTA, citrate and heparin plasma can be used for detection. Centrifuge at 1000 × g for 30 minutes to remove particles.
3. The cell supernatant --- 1000 × g centrifuged for 10 minutes to remove particles and polymers.
4. Storage ------ If the sample is not used immediately, it should be divided into small parts and stored at -70 ℃ to avoid repeated freezing. If possible, do not use hemolysis or hyperlipidemia. If there are a large number of particles in the serum, centrifuge or filter before testing. Do not thaw at 37 ° C or higher. Thaw at room temperature and ensure that the sample is thawed evenly and adequately.

Reagent preparation
1. Standards: Serial dilutions of standards should be prepared during the experiment and cannot be stored. Before dilution, the standard was mixed by shaking. The dilution ratio is as follows:
4.0 g / L (Standard No. 6) The original concentration is added directly to 50ul without dilution.
2.0 g / L (No. 5 standard) 100ul of the original standard is added to 100ul of standard dilution
1.0 g / L (Standard No. 4) Add 100ul of Standard No. 5 to 100ul of Standard Diluent
0.5 g / L (No. 3 standard) 100ul of No. 4 standard is added with 100ul of standard diluent
0.25 g / L (Standard No. 2) Add 100ul of Standard No. 3 to 100ul of Standard Diluent
0.125 g / L (No. 1 standard) 100ul of No. 2 standard is added with 100ul of standard diluent
0 g / L (blank control) The original concentration is directly added to 50ul without dilution.

2. Dilution of washing buffer (50 ×): 50-fold dilution with distilled water.

Steps
1. Before use, mix all reagents thoroughly. Don't make the liquid generate a lot of foam, so as to avoid adding a large number of bubbles during sample addition, which will cause errors in sample addition.
2. The number of slats required is determined by the number of samples to be tested plus the number of standard products. It is recommended to make multiple holes for each standard and blank hole. Each sample is determined according to its own quantity, and those that can use multiple holes can be used as much as possible.
3. Add 50ul of the diluted standard to the reaction well and 50ul of the sample to be tested in the reaction well. Immediately add 50ul of biotin-labeled antibody. Cover the membrane plate, gently shake and mix, and incubate at 37 ° C for 1 hour.
4. Shake off the liquid in the wells, fill each well with washing liquid, shake for 30 seconds, shake off the washing liquid, pat dry with absorbent paper. Repeat this operation 3 times. If washing with a plate washer, the number of washes is increased once.
5. Add 80 ul of streptavidin-HRP to each well, gently shake and mix, and incubate at 37 ° C for 30 minutes.
6. Shake off the liquid in the wells, fill each well with washing liquid, shake for 30 seconds, shake off the washing liquid, pat dry with absorbent paper. Repeat this operation 3 times. If washing with a plate washer, the number of washes is increased once.
7. Add 50ul of substrate A and B to each well, mix gently by shaking, and incubate at 37 ° C for 10 minutes. Avoid light.
8. Remove the enzyme labeling plate and quickly add 50ul of stop solution. After adding the stop solution, the results should be measured immediately.
9. The OD value of each well was measured at a wavelength of 450 nm.

Recommended experimental protocol Standard concentration (g / L)
A 4.0 4.0 sample sample sample sample sample sample sample sample sample sample sample
B 2.0 2.0 sample sample sample sample sample sample sample sample sample sample sample
C 1.0 1.0 sample sample sample sample sample sample sample sample sample sample sample
D 0.5 0.5 sample sample sample sample sample sample sample sample sample sample sample
E 0.25 0.25 sample sample sample sample sample sample sample sample sample sample sample
F 0.125 0.125 sample sample sample sample sample sample sample sample sample sample sample
G 0 0 sample sample sample sample sample sample sample sample sample sample sample
H sample sample sample sample sample sample sample sample sample sample sample sample sample


Limitation
The results of standard No. 6 and above are non-linear, and accurate results cannot be obtained based on this standard curve.

Kit performance
1. Sensitivity: The smallest detection concentration is less than No. 1 standard. Linearity of dilution. The correlation coefficient R between the linear regression of the sample and the expected concentration is 0.990.
2. Specificity: Does not react with other cytokines.
3. Repeatability: The coefficients of variation within and between plates are less than 10%.

Judgment and analysis of results
1. Instrument value: read the OD value of each well on a microplate reader with a wavelength of 450nm

2. Take the absorbance OD value as the ordinate (Y), and the corresponding α1-AGP standard concentration as the abscissa (X), and make the corresponding curve. The α1-AGP content of the sample can be converted from the standard curve according to its OD The corresponding concentration.

3. Detection value range: 0-4.0g / L

4. Sensitivity: 0.01 g / L

Silicone Measuring Cups INDESTRUCTIBLE CONSTRUCTION - FDA approved silicone contains no BPA, can handle extreme high & low temperatures. and will never break.


SILICONE MEASURING CUPS EASY TO CLEAN - Dishwasher safe, but only takes seconds to clean by hand. The measurements are molded into each cup, so unlike our competitors, the ink won't leak into your food.


SILICONE MEASURING CUPS ACCURATE - Set of 4 includes full cup (250mL), half cup (125mL), third cup (80mL), and quarter cup (60mL) sizes, all clearly marked in both metric and US imperial. Easy to read big numbers, as well!


SILICONE MEASURING CUPS ERGONOMIC - Thick silicone handles with dimples make these easy to grip no matter the situation. Plus, the pouring spout makes life a lot less messy.


INNOVATIVE DESIGN - Collapsible, nested design makes these the easiest measuring cups to store that you'll ever own. Flexible silicone makes pouring super easy, while eliminating messes.


SILICONE MEASURING CUPS Details:


Material:100%FDA Silicone measuring cup
Size:60ML,80ML,125ML,250ML
Weight:118g/Set

Color:any pantone color

SILICONE MEASURING CUPS Packaging: 1set per poly bag, 125 sets per carton, carton size:36.5*35*33cm, N.W./G.W.: 10.08kg/ 11kg.


Silicone Measuring Cups

Silicone Measuring Cups,Measuring Cup Set,Small Measuring Cups,Pyrex Measuring Cup

Shenzhen Feiaoda Technology Co.,Ltd , https://www.flysilicone.com