a102 bacterial incubator manual

<

Purpose: To establish standard operating procedures for the cleaning and use of bacterial incubators

Scope of application: use, cleaning and maintenance of bacterial incubator.

Responsibility: The inspectors who use and maintain the bacteria incubator are responsible for the implementation of this regulation, and the director of the laboratory assumes the supervision and inspection responsibility for the effective implementation of this regulation.

program:

1. Clean 0.1% Xin Jie Er Jie or 75% ethanol every week, wipe the dead corners in the box and may be contaminated, and use disinfectant alternately, and perform bacterial inspection once a month.

2. Use procedures

2.1 Before energizing, first check the electrical performance of this box, and pay attention to whether there is a disconnection or leakage.

2.2 Select a suitable mercury thermometer and insert it into the vent hole on the top of the box to observe the temperature inside the box at any time. The temperature in the box should be based on the indication of the thermometer, and cannot be based on the number indicated by the temperature adjustment knob.

2.3 When in use, plug in the power plug, turn the power switch to the "up" position, and start working.

2.4 Temperature adjustment: Turn the temperature adjustment knob clockwise and the red indicator light is on, indicating that the electric heater is heating up and the temperature in the box has started to rise. The red indicator light will automatically turn off, and then fine-tune the green indicator light to make it constant temperature. The desired temperature is reached.

2.5 The average load of the sample shelf is 15kg. When placing the sample, do not over-dense and overload it, so that the air flow is smooth and the temperature in the box is consistent.

2.6 If you want to observe the sample in the working room, you can open the box door and observe from the glass door in the box, but the box door should not be opened often to avoid affecting the constant temperature.

2.7 When using electricity, never touch the electrical parts in the space on the left side of the box with water, or wipe with a damp cloth and rinse with water.

2.8 After each use, fill in the use registration.

3. Matters needing attention

3.1 The box should be placed in a dry and level room without using other fixing devices.

3.2 When the device fails, you must first cut off the power, remove the side door, and check them one by one according to the circuit diagram.

3.3 This box is a non-explosion-proof incubator, so do not put flammable or volatile items in the drying box to avoid accidents.

2.4 An iron shell knife should be installed in the power supply line for this box only.

4.5 When the equipment is not in use, the external power supply should be cut off.

3.6 When cultivating, the top air hole should be unscrewed properly to allow the humid air in the box to escape.

3.7 If the temperature inside the box is suddenly high or low, mostly due to the failure or sticking of the contact of the temperature controller, the side chamber should be opened and wiped. If the temperature is still not constant, it should be further repaired or replaced temperature control components or equipment.

From: http: //

Propane Gas Stove And Burner Accessories. On holidays, you can invite your friends to have a party outdoors. You can use the Adjustable Camping Burner Stove, Adjustable Outdoors Burner Stove, Super Propane Gas Cooking Cooker, etc. It would must make your party more funny. At ARC, safety & heat go hand in hand. Don`t wait! Just buy it now!